3.2 Remote control commands

795 KFT Titrino

78

Reports which are sent from the Titrino are marked with space (ASCII 32) and

' at the beginning. Then an individual identifier for each report follows. Reports

which are triggered by RS232 ($G) have the same introducer but without pre-

ceding space, i.e. they start with '.

3.2.2.61. Info.Checksums $G

Info.Checksums.MPList read

only

Info.Checksums.ActualMethod read

only

The checksums can be used to identify the content of a file unequivocally,

e.g. files with identical content

have identical results of the checksums. An empty file has checksum "0". The

calculation of the checksums is triggered with $G.

.MPList: Result of the checksum of the current measuring point list.

.ActualMethod: Result of the checksum of the current method in the working

memory. Identical methods with different method names have the

same results of the checksum.

3.2.2.62. Info.DetermData $G

Info.DetermData.Write ON,

OFF

Info.DetermData.MPList.1.Attribute read

only/read+write

Info.DetermData.ExV read

only/read+write

Info.DetermData.MPList.1.X read

only/read+write

Info.DetermData.MPList.1.Y read

only/read+write

Info.DetermData.MPList.1.Z1 read

only/read+write

Info.DetermData.MPList.1.Z2 read

only/read+write

for every measuring point

Determination data in hexadecimal format. A measuring point list is available.

Recalculation of the measuring data is triggered with $G.

.Write:

With "ON", the following nodes can be overwritten:

&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),

and &Mode.Name.

.ExV:

Volume of the exchange unit, with which the determination was

executed

.MPList.1.Attribute: Attribute

.MPList.X: X coordinate, time

.MPList.Y: Y coordinate, volume

.MPList.Z1: Z1 coordinate, measuring value

.MPList.Z2: without meaning