795 KFT Titrino

76

3.2.2.52. SmplData.OFFSilo.Id1

up to 8 ASCII characters

SmplData.OFFSilo.Id2

up to 8 ASCII characters

SmplData.OFFSilo.Id3

up to 8 ASCII characters

SmplData.OFFSilo.ValSmpl

6-digits, sign and decimal point

SmplData.OFFSilo.UnitSmpl

up to 5 ASCII characters

Current sample data.

The identifications Id1...Id3 can be used in formulas as sample-specific calcu-

lation constants C21...C23.

If "no unit" is desired for the unit of the sample size, the blank string must be

entered.

3.2.2.53. SmplData.ONSilo.Counter.MaxLines read

only

SmplData.ONSilo.Counter.FirstLine read

only

SmplData.ONSilo.Counter.LastLine read

only

Information on silo memory.

.MaxLines: Maximum possible number of silo lines.

.FirstLine: Lowest valid silo line.

.LastLine: Last occupied silo line.

3.2.2.54. SmplData.ONSilo.EditLine.1.Method

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.Id1

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.Id2

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.Id3

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.ValSmpl 6-digits, sign and dec.point

SmplData.ONSilo.EditLine.1.UnitSmpl

up to 5 ASCII characters

SmplData.ONSilo.EditLine.1.C24 read

only

SmplData.ONSilo.EditLine.1.C25 read

only

SmplData.ONSilo.EditLine.1.Mark read

only

etc., up to .255

Contents of a silo line.

.Method: Method used to process the sample, from the method memory or

from the card.

.Id:

The identifications Id1...Id3 can also be used as sample-specific

calculation constants C21...C23 in formulas.

.UnitSmpl: If "no unit" is desired for the sample size, the blank string must be

entered.

.C24, .C25: Results which have been assigned to C24 and C25.
.Mark:

Mark of the silo line: "∗"=deleted line, "+"=line which is worked

off, "–"= line which is worked off and not valid for silo calcula-

tions (deleted), "/" last worked-off line, where recalculation can

still be done. Silo lines which have been worked off are "read

only".

3.2.2.55. SmplData.ONSilo.DelLine $G

SmplData.ONSilo.DelLine.LineNum 1...255, OFF

Deletion of a silo line. The line # is deleted with &SmplData.ONSilo.DelLine

$G. If a formerly deleted line is edited again, it becomes valid (function "unde-

lete").