Metrohm 795 KFT Titrino User Manual

Page 55

Advertising
background image

3.1 General rules

795 KFT Titrino

51

E196

Result is out of limits.

Exit: The error message disappears on next start or on recalculation.

E197

Sample size is out of limits.

Exit: The error message disappears on next start or on introduction of new

sample size.

E198

Validation interval is expired.

Exit: The error message disappears on next start or clear counter with

&Config.Monitoring.Validation.ClearCount $G.

E199

Service date is reached.

Exit: The error message disappears on next start or change date in

&Config.Monitoring.Service.Date.

E203

No Oven parameters: Oven not (correctly) connected.

Exit: The error message disappears on next start. If you don't wish oven

parameters in your report, select &Mode.Parameter.Presel.Oven "no" in your

method(s).

E212

Transmission error from Remote Box. Unknown characters.

Exit: Rectify error and switch Titrino off and on again.

E213

Time-out error from PC keyboard (Remote Box)

Exit: Rectify error and switch Titrino off and on again.

E214

Check Remote Box. Remote Box not (properly) connected but activated in

&Config.Periph.RemoteBox.

Exit: Rectify error and switch Titrino off and on again.