Metrohm 795 KFT Titrino User Manual

Page 83

Advertising
background image

3.2 Remote control commands

795 KFT Titrino

79

3.2.2.63. Info.TitrResults.RS.1.Value read

only

etc., up to .9

Info.TitrResults.EP.1.V read

only

Info.TitrResults.EP.1.Meas read

only

etc., up to .2

Info.TitrResults.Var.C40 read

only/read+write

etc., up to .C45

Info.TitrResults.Var.DTime read

only/read+write

.RS: Values of the calculated results.

.EP: Endpoints:

Volume coordinate in mL, e.g. "1.2340"

Measured value coordinate in mV (with Ipol) "–241" or µA (with

Upol) "43.7".

.Var: Various variables. You may overwrite the variables C40...C45, see

3.2.2.62.

C40: Initial measured value in mV (with Ipol) "41" or µA (with

Upol).

C41: End volume in mL, "12.5360".

C42: Time from start of titration to end in s, "62".

C43: Volume drift on start of KFT titration from the conditioning in
µL/min, "3.5".

C44: Temperature in °C.

C45: Start volume in mL, "2.800".

DTime: Time for the drift correction in KFT with conditioning.

3.2.2.64. Info.StatisticsVal.ActN read

only

Info.Statistics.1.Mean read

only

Info.Statistics.1.Std read

only

Info.Statistics.1.RelStd read

only

etc. up to .9

The current values of the statistics calculation.

$Q sends, e.g.

ActN: Current value of the individual results

"3"

Data for MN1:

Mean: Mean value (decimal places as in result)

"3.421"

Std: Standard deviation (1 decimal place more than in result)

"0.0231"

RelStd: Relative standard deviation (in %, 2 decimal places)

"0.14"

3.2.2.65. Info.SiloCalc.C24.Name read

only

Info.SiloCalc.C24.Value read

only

Info.SiloCalc.C24.Unit read

only

for

.C25 as for .C24

Info.SiloCalc.C26.ActN read

only

Info.SiloCalc.C26.Mean read

only

Info.SiloCalc.C26.Std read

only

Info.SiloCalc.C26.RelStd read

only

for

.C27 as for .C26

The current values from the silo calculations. C26 is the mean value out of the

C24 variables; C27 comes from C25.