Metrohm 795 KFT Titrino User Manual
Page 61
Advertising
3.2 Remote control commands
795 KFT Titrino
57
"Configuration",
continuation
├
.RSSet1
Settings RS232, 1
$G 3.2.2.49.
│ ├
.Baud Baud
rate
300,600,1200,2400,4800,
│
│
9600,19200,38400,57600,
│
│
115200
ditto
│ ├
.DataBit
Number of data bits
7, 8 ditto
│ ├
.StopBit
Number of stop bits
1, 2
ditto
│ ├
.Parity Parity
even,
odd,
none
ditto
│ ├
.Handsh Handshake
HWs, SWchar,
│
│
SWline, none
ditto
├
.RSSet2
as for RS1
│
├
.ComVar
Values of common variables
│ ├
.C30 C30
0... ±999999
3.2.2.50.
│ ├
up to C39
0... ±999999
Advertising