3.5 Transport locking screws

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

872 Extension Module IC Module

3

Drainage tubing connection
For supplying escaped liquid through the
connected drainage tubing to the leak sen-
sor.

4

Drainage tubing connector
For draining escaped liquid from the base
tray through the connected drainage tubing.

5

Drainage tubing
Section of the 6.1816.020 silicon tubing.
Guides escaped liquid into a waste con-
tainer.

Plug the drainage tubing (10-2) onto the drainage tubing connec-
tor (10-1) of the bottle holder and shorten to required length.

Plug the other end of the drainage tubing (10-2) onto the drain-
age tubing connector (10-3) of the base tray.

Plug the drainage tubing (10-5) onto the drainage tubing connec-
tor (10-4) and guide the other end into a waste container.

3.5

Transport locking screws

To avoid damage to the high pressure pump and vacuum pump during
transport, the pumps are secured with transport locking screws .

Remove these transport locking screws before the initial start-up.

Removing transport locking screws

1 Remove all of the transport locking screws with the 6.2621.030 4

mm hexagon key and keep them in a safe place.

Warning

In order to avoid damage to the pump, the transport locking screws
must be remounted each time the instrument undergoes major trans-
port.