7 Troubleshooting
872 Extension Module IC Module
■■■■■■■■
59
Problem
Cause
Remedy
Eluent – Blockage in the
eluent path.
Check the eluent path
Precision problems -
significant scatter-
ing of the measured
values
Injection valve – sample
loop.
Check installation of the sample loop (see
Chapter 4.9.1, page 35).
Sample – rinsing volume
too low.
Increase rinsing time (see Chapter 6.7, page
55).
Injection valve – defective.
Request Metrohm Service.
Unexpected change
to the retention
times in the chroma-
tograms
Eluent – Gas bubbles in the
eluent.
Check connections of the eluent degasser (see
Chapter 4.5, page 27).
High pressure pump –
defective.
Request Metrohm Service.
Vacuum is not being
built
Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.
■
Seal the connector Vacuum tightly with a
6.1446.040 threaded stopper.