Figure 17, Eluent degasser – Metrohm 872 Extension Module User Manual
Page 36
4.5 Eluent degasser
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
872 Extension Module IC Module
Connecting the eluent degasser
5
6
1
2
3
4
Figure 17
Eluent degasser
1
Eluent degasser input
2
Eluent degasser output
3
Tubing flare
With tubing nipple.
4
Clamping screw
5
Eluent aspiration tubing (6.1834.080)
For aspirating the eluent. The clamping
screw (17-
) is firmly mounted.
6
Connection tubing (6.1834.090)
Connection from the eluent degasser to the
high pressure pump (see Chapter 4.6, page
29). The clamping screw (17-
mounted.
1
Caution
The clamping screws (17-4) must be tightened carefully. Use the
wrench (6.2621.050) to do this.
■
Insert the eluent aspiration tubing (17-5) into the eluent degasser
input (17-1).
■
Carefully tighten the clamping screw (17-4).
2
■
Insert connection capillary (17-6) (the end with the longer clamp-
ing screw (17-4)) into the eluent degasser output (17-2).
■
Carefully tighten the clamping screw (17-4).
■
Connect the other end of the connection capillary (17-6) (with
the shorter clamping screw ) to the high pressure pump (18-9)
(see "Connecting inlet to the high pressure pump", page 30).