4.5 Eluent degasser

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

872 Extension Module IC Module

Connecting the eluent degasser

5

6

1

2

3

4

Figure 17

Eluent degasser

1

Eluent degasser input

2

Eluent degasser output

3

Tubing flare
With tubing nipple.

4

Clamping screw

5

Eluent aspiration tubing (6.1834.080)
For aspirating the eluent. The clamping
screw (17-

4

) is firmly mounted.

6

Connection tubing (6.1834.090)
Connection from the eluent degasser to the
high pressure pump (see Chapter 4.6, page
29)
. The clamping screw
(17-

4

) is firmly

mounted.

1

Caution

The clamping screws (17-4) must be tightened carefully. Use the
wrench (6.2621.050) to do this.

Insert the eluent aspiration tubing (17-5) into the eluent degasser
input (17-1).

Carefully tighten the clamping screw (17-4).

2

Insert connection capillary (17-6) (the end with the longer clamp-
ing screw (17-4)) into the eluent degasser output (17-2).

Carefully tighten the clamping screw (17-4).

Connect the other end of the connection capillary (17-6) (with
the shorter clamping screw ) to the high pressure pump (18-9)
(see "Connecting inlet to the high pressure pump", page 30)
.