7.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

58

■■■■■■■■

872 Extension Module IC Module

7 Troubleshooting

7.1

Problems and their solutions

Problem

Cause

Remedy

Marked rise in pres-
sure

6.2821.120 inline filter
blocked.

Replace 6.2821.130 filter (see Chapter 6.5,
page 53)
.

Injection valve – valve
blocked.

Have the valve cleaned (by Metrohm service
technicians).

Drift of the baseline

Eluent – Evaporation of the
organic solvent in the elu-
ent.

Check the eluent bottle attachment (see Figure
14, page 25)
.

Peak areas lower
than expected

Sample – leak in the sam-
ple path.

Check the sample path.

Sample – blockage in the
sample path.

Check the sample path.

Sample – sample loop not
(completely) filled.

Prolong the sample transfer time.

Very noisy baseline

High pressure pump – con-
taminated pump valves.

Clean pump valves (see Chapter 6.4.2, page
43)
.

Eluent – Leak in the eluent
path.

Check the eluent path

Eluent – Blockage in the
eluent path.

Check the eluent path

High pressure pump –
defective piston seals.

Replace piston seals (see Chapter 6.4.2, page
43)
.

Pulsation damper not con-
nected.

Connect pulsation damper (see Chapter 4.8,
page 34)
.

Background conduc-
tivity too high

Incorrect eluent.

Change eluent (see Chapter 6.3.2.3, page 42).

Individual peaks
greater than expec-
ted

Sample – carry-over of the
samples from previous
measurement.

Rinse system longer between two samples.

Poor reproducibility
of the retention
times

Eluent – Leak in the eluent
path.

Check the eluent path