8.7 Injection valve

■■■■■■■■■■■■■■■■■■■■■■

62

■■■■■■■■

872 Extension Module IC Module

Safety shutdown

Function

Automatic shutdown upon reaching the pressure limit values

Maximum pres-
sure limit

Adjustable from 0.1…50 MPa (1…500 bar)

The pump is automatically shut down at the first piston stroke
above the maximum limit value

Minimum pres-
sure limit

Adjustable from 0…49 MPa (0…490 bar)

The shutdown mechanism is inactive at 0 MPa

The shutdown mechanism only becomes active 2 minutes after sys-
tem start

The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit

Gradient capacity

Isocratic or gradient (extendable to quaternary)

Profile

Step, linear, convex and concave

Resolution

< 1 nL/min flow increments

8.7

Injection valve

Actuator time

typ.100 ms

Max. operating
pressure

35 MPa (350 bar)

Material

PEEK

8.8

Interfaces

Auxiliary

1 DSUB plug 15-pin (female)

Analog Output

Analog Output (optional)

8.9

Safety specification

Design / testing

EN/IEC 61010-1

UL 61010-1

CSA-C22.2 No. 61010-1

Protection class III