background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

872 Extension Module IC Module

■■■■■■■■

21

4 Installation

4.1

About this chapter

The Installation chapter contains

this overview

a brief set of instructions for the installation of the 872 Extension Mod-
ule IC Module (see Chapter 4.2, page 21). At each step you will find
cross-references to more detailed installation instructions for individual
components, should you require such aids.

an installation diagram (see Chapter 4.3, page 22), showing a com-
pletely installed instrument.

several chapters with detailed installation instructions for all compo-
nents, including those that are already installed at the time the instru-
ment is delivered.

4.2

Installation overview

Note

A number of the capillary connections are already connected at the
time the instrument is delivered.

The following work steps must still be carried out:

Installing 872 Extension Module IC Module

1 Installing the eluent path

Tighten the clamping screw of the 6.1834.080 eluent aspiration
tubing to the eluent degasser input.
Equip the free end of the eluent aspiration tubing (11-1) and con-
nect it to the eluent bottle (see Chapter 4.4, page 23).

Tighten one end of the 6.1831.010 PEEK capillary with a
6.2744.014 PEEK pressure screw to the connector 4 of the injec-
tion valve (see Chapter 4.9.1, page 35).
Shorten to required length with the capillary cutter and with the
aid of a 6.2744.040 coupling and two 6.2744.010 pressure
screws connect to the column outlet capillary in the IC instru-
ment.