Fine-tuning meter pressurization, Alternate skip/double check, Seed delivery fan adjustments – Great Plains YP2425A Operator Manual User Manual

Page 65: Seed delivery butterfly valves, The “ alternate skip/double check ” belo

Advertising
background image

Great Plains Manufacturing, Inc.

Table of Contents

Index

Adjustments

61

2014-02-10

Table of Contents

Index

401-626M

Fine-Tuning Meter Pressurization

After several passes, and you are comfortable with the
planter’s operation, you can optimize meter
pressurization. Use this method during longer passes, so
there are fewer distractions (and see sidebar regarding
Milo).

16. Observe the current overall average population

reported by the monitor.

17. Adjust the meter pressurization down, in discrete

steps, waiting 5-10 seconds between adjustments,
until skips occur (actual population begins falling
below target). Note the pressure at which skips
begin.

18. Restore pressure to the initial value at step 16.

19. Adjust the meter pressurization up, by periodic small

increments, until doubles occur (actual population
begins rising above target). Note the pressure at
which doubles begin.

Note: If unable to adjust up to doubles (or skips happen

instead), seed delivery is probably being starved of
air. Increase fan rpm and repeat limit search.

20. Adjust meter pressurization to a value halfway

between the limits established at step 17 and
step 19.
This is the ideal value for your crop, disk and
population, providing equal margin against skips and
doubles. Record this value for future use.

Seed Delivery Fan Adjustments

Refer to Figure 72
The primary control for seed delivery air is fan rpm. The
servo vane (not shown), under the control of the split air
system, takes whatever is needed to maintain meter
pressurization (the bottom three manifold outlets

).

Remaining air is sent to the top outlets for seed delivery.

There are two butterfly valves at the top fan outlets:
left air box

, and right air box

. The recommended

butterfly valve setting is 0

°. Recommended initial fan

speed depends on planter configuration. See page 44.

Adjust the basic fan rate with the tractor hydraulic system
and the fan rpm display of the seed monitor. Do not run
at speeds over 4000 rpm or speeds under 2000 rpm.

Seed Delivery Butterfly Valves
These valves are factory pre-set to the suggested 0

°

(wide open).

Unusual situations can cause an imbalance of seed flow
between left and rights side of the planter, such as
dissimilar bulk seed boxes on each side.

If meters on one side of the planter are being starved of
seed, first try increasing the fan rpm. If the fan rpm is
already high, or increasing it causes other problems, use
the valves.

In these cases, use one butterfly valve to reduce the air
flow to one air box. Leave the valve on the other side at
zero.

Set the valve on the high-flow side. Start with a setting in
the 20-30

° range. Re-prime all meters (page 88). make

further valve adjustments until you achieve a balanced
airflow.

Alternate Skip/Double Check
Small seeds, particularly Milo, are less reliably sensed as
skips or doubles by the seed tube sensors, and watching
monitor population is unlikely to locate the pressure
limits. This test can also be quicker for all seeds.

A. After planting a few passes with initial settings,

remove the rain covers from several rows (use rows
with a variety of seed hose lengths and routes).

B. Make a meter pressurization adjustment. Resume

planting for a pass or less.

C. Stop planter motion but leave fan running.

D. Inspect the seed disks closely. Look for empty seed

cells (skips) and cells with multiple seeds (doubles).

E. Repeat step B-step D until limits are established.

Record limits. Re-install rain covers. Plant with
median settings.

Figure 72

Fan Butterfly Valves

29773

3

2

U

D

F

B

L

R

1

1

2

3

Advertising