Spring adjustment – Amico Falcon Combo Arm User Manual

Page 36

Advertising
background image

36 Amico Accessories Inc.

Spring Adjustment

WARNING

Swivel-Post interior contains sharp edges, wear protective equiment when performing the tasks

below.

1. The spring in the Swivel-Post can be replaced if the monitor weight changes in order to maintain the appropriate

counterbalancing.

2. Remove the cap on top of the Swivel-Post to reveal the two screws. Use a philips head screwdriver (#2) to

remove the two screws (Figure 1). Also remove the Swivel-Post cover.

3. Remove the top plate and slide out the VESA Head to reveal the gas spring underneath. Replace the gas spring

with the new gas spring, ensure that there is a HEX nut on the thin side of the replacement spring and the nut is

facing up (Figure 2).

4. Replace the VESA Head, re-attach the top plate with the screws and replace the cap and the swivel post cover.

(Figure 2)

HEX nut

Gas Spring

(Figure 1)

Swivel-Post

VESA plate

Screws

Swivel-Post Cover

Top Cap

Top Plate

SECTION 4: Adjustments

Advertising