Spring adjustment – Amico Falcon Combo Arm User Manual
Page 36
36 Amico Accessories Inc.
Spring Adjustment
WARNING
Swivel-Post interior contains sharp edges, wear protective equiment when performing the tasks
below.
1. The spring in the Swivel-Post can be replaced if the monitor weight changes in order to maintain the appropriate
counterbalancing.
2. Remove the cap on top of the Swivel-Post to reveal the two screws. Use a philips head screwdriver (#2) to
remove the two screws (Figure 1). Also remove the Swivel-Post cover.
3. Remove the top plate and slide out the VESA Head to reveal the gas spring underneath. Replace the gas spring
with the new gas spring, ensure that there is a HEX nut on the thin side of the replacement spring and the nut is
facing up (Figure 2).
4. Replace the VESA Head, re-attach the top plate with the screws and replace the cap and the swivel post cover.
(Figure 2)
HEX nut
Gas Spring
(Figure 1)
Swivel-Post
VESA plate
Screws
Swivel-Post Cover
Top Cap
Top Plate
SECTION 4: Adjustments