Mounting to the mrs (monitor rail system) – Amico Falcon Combo Arm User Manual

Page 8

Advertising
background image

8 Amico Accessories Inc.

Mounting To The MRS (Monitor Rail System)

FIT Station

WARNING

DO NOT attempt to mount/remove the FIT Station from the rail when loaded with any

instrument

WARNING

DO NOT mount the FIT Station anywhere near or above a Patient’s head.

WARNING

To prevent the Mounting Adapter from sliding down the MRS, ensure the four set screws

are fastened to the tightest possible position, serious injuries could result if the set screws are not properly
fastened. (Figure 1)

WARNING

Ensure rail is stable and properly installed before mounting FIT Station.

NOTE

The top and bottom stopper are there to prevent the Mounting Adapter from sliding out of the

MRS. The Mounting Adapter must be taken out from the top of the MRS with the top stopper pressed back into
the MRS (Figure 2)

1. Install Height locking Lever (See Page 25)

2. Engage the Height Locking Lever with the AHM in the highest position.

3. Lift up the FIT Station and guide the Mounting Adapter into the top of the MRS Rail (Figure 1).

4. Once the arm is positioned at the desired height in the MRS, tighten all four set screws to the tightest possible

position on the Mounting Adapter using a 1/8’’ HEX key.

(Figure 2)

Bottom
Stopper

Top Stopper

MRS Rail: Length for representational use only

(Figure 1)

MRS

Mounting
Adapter

1/4-20 Set
Screws

SECTION 2: Installation on Rail Systems