CPU Mount

WARNING

Ensure CPU is within weight and size rating of the CPU mount. If not, please contact Amico

Accessories at (1-877-264-2697)

WARNING

The CPU mount can be mounted anywhere at any height on the rail. However, DO NOT mount

anywhere above a patient’s head.

WARNING

Ensure rail is stable and properly installed before mounting FIT Station.

1. Determine a suitable location on the MRS for the CPU to make sure it isn’t above a patient’s head.

2. Lower the CPU mount down the MRS to the desired height. Tighten the two set screws in the Mounting Adapter

to the tightest possible position using a 1/8” HEX key. (Figure 1)

3. Mount the CPU and secure it by tightening the two thumb knobs.

(Figure 1)

Thumb
Knobs

CPU
Mount

MRS

1/4-20 Set Screws on
Mounting Adapter

CPU slides
into slot

OSHPD Anchor Pre-Approval OPA-1743

In compliance with standards established for the anchor and installation of instrument support systems by the
California Office of Statewide Health Planning and Development (OSHPD), Amico Accessories Inc. has obtained pre-
approval for MRS 19’’ (48.26cm) Wall Channels or longer with extension Arm. For proper installation requirements
please refer to our OPA Document.

SECTION 2: Installation on Rail Systems