assembly. Figure 5-10 shows the panel cutout dimensions, and the dimensions for the panel thickness. It requires that
the following already be done:

1. Your line voltage rating has been checked against the meter rating on the Product ID and serial number label on the

meter case. See Section 5.2.1.

2. You have configured all jumpers-those on the main board as well as those on any optional boards. See Section 5.2.2

for main board jumper configuration and the appropriate sections for optional board jumper configuration.

3. You have installed all optional boards and inserted the main board assembly back into the case. See Section 5.2.3.
4. You have wired P1, the AC power connector, and P2 the input output control connector; connectors are not installed

in the meter, but are ready to be installed. See Section 7.

5. You have wired all connectors for optional boards; connectors are not connected to the meter, but are ready to be

installed.


If all of these steps are done, use Figure 5-10 as a guide:

CAUTION: Connectors with the wiring connections will be installed after mounting the unit.

Figure 5-10. Panel Mounting Assembly


6. Punch or cut a hole in the panel using the panel cutout dimensions in Figure 5-10. Remove burrs and paint the panel

as required.

7. Insert the panel-mount gasket around the rear of the case and slide it forward to the bezel (if it

’s not already in place).

8. Working from the front of the panel, insert the case assembly, rear end first, all the way into the panel cutout such that

the gasket firmly backs the panel surface.

9. Working from the rear of the panel, slide the sleeve forward over the case and up to the panel surface.

The panel should now be sandwiched between the bezel-backed gasket in front and the sleeve in back.

10. Replace the thumbnuts that secure the sleeve tabs to the case.

WARNING: Do not “turn-on” the ac power and input signals until all connections are connected to the meter.

CF 156