Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual

Page 53

Advertising
background image

This scaling method based on 2 Input values entered with 2 corresponding readings. Up to 10 Linearization Points can be
selected to customize the transducer curve.

MP.SC.OF

Enter MULTI-POINT SCALE & OFFSET MENU:

Press ‘MENU

1) Four times or scroll though Main Menu items, select and enter "MP.SC.OF".

Press ‘RESET/ENTER

2) Display shows "NUM.PNt" Number of Points menu.

Press ‘RESET/ENTER

3) Display flashes some 2-digit number or previous setting which represents the number
of Scale & Offset points.

Press ‘▲/MAX

4) Scroll through "00" to "09" for your choice of Scaling method that allows users to
select up to 10 points for transducer curve’s linearization purposes.

Press ‘RESET/ENTER

5) Display shows "StOREd" momentarily then advances to "OUt.CNF", if "00" was
selected. Otherwise displays will flash either of the following 2 options: "ON.L INE" or
"MANUAL". Select your choice for Scaling method.

ON.L INE

IF "ON.L INE" and "NUM.PNt" = 01 are YOUR CHOICE:
(On-Line Calibration with 2 points)

Press ‘RESET/ENTER

6) Display stops flashing "ON.L INE" On-line Calibration or Scaling.

Press ‘RESET/ENTER

7) Display shows "INPUt 0".

Press ‘RESET/ENTER

8) Display shows the actual live-input signal being received and equal to approximately
the minimum input signal or 0% of known load range.

Press ‘RESET/ENTER

9) Display shows "StOREd" momentarily and advances to "REAd 0".

Press ‘RESET/ENTER

10) Display shows last stored value of "REAd 0" with 1st digit flashing.

Press ‘▲/MAX’ & ‘►/MIN’ 11) To enter the value of "REAd 0".
Press ‘RESET/ENTER

This value corresponds to "INPUt 0" in terms of some meaningful engineering units. To
show "INPUt 0" as minimum or 0% as enter "REAd 0" value = 000000 (Example)

Press ‘RESET/ENTER

12) Display shows "INPUt 1".

Press ‘RESET/ENTER

13) Display shows the 2nd higher live-input signal being connected for your preference
as equal to approximately 100% of the transducer range or the max. Input signal.

Press ‘RESET/ENTER

14) Display shows "StOREd" momentarily and advances to "REAd 1".

Press ‘RESET/ENTER

15) Display shows last stored value of "REAd 1" with 1st digit flashing.

Press ‘▲/MAX’ & ‘►/MIN’ 16) To enter the value of "REAd 1". This value corresponds to "INPUt 1" in term of

some meaningful engineering units. To show "INPUt 1" as maximum or 100% as enter
"REAd 1" value = 000100 (example).

Press ‘RESET/ENTER

17) Display shows "StOREd" momentarily and advances to "OUt.CNt" Output
configuration menu.

MANUAL

IF "MANUAL" and "NUM.PNt"=01 are YOUR CHOICE
(Scaling without Known Loads or based on specifications)

CF 156

48

M2544/N/0505