Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual

Page 50

Advertising
background image

ON.L INE
Press ‘RESET/ENTER

30) Display flashes "ON.L INE" On line method.

Press ‘▲/MAX

31) Scroll between 2 available selections: "ON.L INE" or "MANUAL" for the configuration
option of your choice. If your choice of selection is "MANUAL", skip to Step 40).

Press ‘RESET/ENTER

32) Display stops flashing and shows "ON.L INE".

Press ‘RESET/ENTER

33) Display shows "INPUt 1" Item #1 of coordinate #1.
Pressing ‘RESET/ENTER’ to sequence through

"INPUt 1", Input 1 ‘s value
"REAd 1", Read 1 ‘s value
"INPUt 2", Input 2 ‘s value
"REAd 2", Read 2 ‘s value

headings for the desired selection to enter new value.

Press ‘RESET/ENTER

34) Display shows "xxxxxx" some 6-digit number, represents the value of "INPUt 1" as
taken from live input source measurement for desired value.

Press ‘RESET/ENTER

35) Display shows "REAd 1" Item #2 of coordinate #1.

Press ‘RESET/ENTER

36) Display shows "xxxxxx" some 6-digit number with first digit flashing, represents the
value of "REAd 1" item #2 of coordinate #1.

Press ‘▲/MAX’ & ‘►/MIN’ 37) Use ‘MAX’ and ‘MIN’ buttons to enter new value of "REAd 1".
Press ‘RESET/ENTER

38) Display shows "INPUt 2" Item #1 of coordinate #2. Continue to proceed with the
selection and value entry for "INPUt 1" and "REAd 1" for "INPUt 2" and "REAd 2", (item
#1 and #2 of coordinate #2) similar to steps above.

Press ‘RESET/ENTER

39) Display advances to "MP.SC.OF" Multi-Scale & Offset configuration menu. If "REAd 1"
item #2 of coordinate #2 was prompted or modified, the controller will also show "StOREd"
stored message momentarily.

MANUAL

Press ‘RESET/ENTER’ 40)

"MANUAL" heading stops flashing.

Press ‘RESET/ENTER

41) Display shows "INPUt 1" Item #1 of coordinate #1.
Pressing ‘RESET/ENTER’ to sequence through

"INPUt 1", Input 1 ‘s value
"REAd 1", Read 1 ‘s value
"INPUt 2", Input 2 ‘s value
"REAd 2", Read 2 ‘s value

headings for the desired selection and continue using ‘MAX’ and ‘MIN’ buttons to manually
enter or modify the values as described the below section.

Press ‘RESET/ENTER

42) Display shows "xxxxxx" some 6-digit number with first digit flashing, represents the
value of "INPUt 1" item #1 of coordinate #1.

Press ‘▲/MAX’ & ‘►/MIN’ 43) Use ‘MAX’ and ‘MIN’ buttons to enter new value of "INPUt 1" item #1 of coordinate #1.
Press ‘RESET/ENTER

44) Display shows "REAd 1". Continue with the sequential items and use ‘MAX’ & ‘MIN
buttons to select and enter values.

Press ‘RESET/ENTER

45) Display advances to "MP.SC.OF" Multi-point Scale & Offset Configuration Menu. If
"REAd 2" item #2 of coordinate #2 was prompted or modified, the controller will also
shows "StOREd" stored message momentarily.


SCALING YOUR METER
Scaling your meter without a sensor or a signal source connected is easily accomplished by using one of the calculated
scale factors as shown in Table below.

Scaling to Display Engineering Units
These calculated scale factors are the numbers that your meter would display if you connected a signal source.

You must complete the procedure for preparing your meter for Scaling as described in Section 16.4 "IN.SC.OF" INPUT
SCALE & OFFSET, "MANUAL". Failure to complete this procedure will result in erroneous readings.

Input Signal Range
Scale Input 1 Value
Scale Input 2 Value

0-20mA

000000
020000

4-20mA

000000
020000

±50mV
-50000

050000

±500mV

-50000

050000

0-20mV

000000
020000

0-30mV

000000
030000

CF 156

45

M2544/N/0505