Cooper Instruments & Systems DFI INFINITY B Ultra High Performance Digital Force Indicator User Manual

Page 54

Advertising
background image

To scale "MANUAL", calculate inputs based on transducer specifications and manually enter them via on the front panel
push buttons. The following example assumes load cells with this specification:

Maximum Load:....................................................................100.0 lb
Output ................................................................................3.0 mV/V
Sensor Excitation ......................................................................10 V
Maximum Sensor Output = 3.0(mV/V) x 10(V) = 30mV


Determine the correct values for Inputs ("INPUt 0" and "INPUt 1"). Calculate "INPUt 0" and "INPUt 1" using the following
equation:

"INPUt" = (Sensor Output) x (Conversion Number)

Conversion number is a coefficient of conversion between input values and real full display range (100000
counts). See table below for proper conversion number.

INPUT RANGE

CONVERSION NUMBER

0 - 100 mV
0 - 1 V
0 - 10 V
0

– 20 mA

100000/(100 x 1)=1000cts/mV
100000/(1000 x 1)= 100cts/V
00000/(1000 x 10)= 10cts/V
20000/ (20 x 1) = 1000cts/mA

18-1. Conversion Number Table


Determine "INPUt 0" and "INPUt 1" Input range. For our transducer select 0 - 100 mV range.

"INPUt 0" = 0 (mV) x 1000(cts/mV) = 000000
"INPUt 1" = 30 (mV) x 1000(cts/mV) = 030000


1. Determine correct values for Display Reading ("REAd 0" and "REAd 1")
2. In most cases,"REAd 0" and "REAd 1" are equal to the minimum and the maximum of the transducer output range.

"REAd 0" = 000000
"REAd 1" = 000100

3. Scaling the controller.

Press ‘RESET/ENTER

18) Display stops flashing "MANUAL" Manual Scaling method menu.

Press ‘RESET/ENTER

19) Display shows "StOREd" momentarily and advances to "INPUt 0" heading.

Press ‘RESET/ENTER

20) Display shows last stored "INPUt 0" value with 1st digit flashing.

Press ‘▲/MAX’ & ‘►/MIN’ 21) To enter "INPUt 0" value (000000)
Press ‘RESET/ENTER

22) Display advances to "REAd 0" heading.

Press ‘RESET/ENTER

23) Display shows last stored "REAd 0" value with 1st digit flashing.

Press ‘▲/MAX’ & ‘►/MIN’ 24) To enter "REAd 0" value (000000).
Press ‘RESET/ENTER

25) Display shows "INPUt 1" heading.

Press ‘RESET/ENTER

26) Display shows last stored "INPUt 1" value with 1st digit flashing.

Press ‘▲/MAX’ & ‘►/MIN’ 27) To enter "INPUt 1" value (030000).
Press ‘RESET/ENTER

28) Display advances to "REAd 1" heading.

Press ‘RESET/ENTER

29) Display shows last stored "REAd 1" value with 1st digit flashing.

Press ‘▲/MAX’ & ‘►/MIN’ 30) To enter "REAd 0" value (000100).
Press ‘RESET/ENTER

31) Display shows "StOREd" momentarily and advances to "OUt.CNF" Output
Configuration Menu.

INPUt value = Xin

RS = Reading Scale

RO = Reading Offset

Display Reading Value = Yrd

IS = Input Scale

IO = Input Offset

( . ) : Multiplication

MS = Multi Scale

MO = Multi Offset

CF 156

49

M2544/N/0505