Operation

OPERATOR’S MANUAL

CenSlide 2000 Urinalysis System

Page 2 - 3

4. If stain is required, add it to the

sample after the chemistry test
strip has been removed and
read (Figure 2.2).

5. Place the cap on the CenSlide

tube and push down firmly until it
clicks and/or its rim is flush with
the top of the CenSlide tube.

6. Holding the CenSlide tube just

below the cap, align it so that the
flat side of the CenSlide tube
conforms to the flat side of the
slot in the rotor head (Figure 2.3).
One to six CenSlide tubes can
be placed into the rotor with
appropriate counter-balancing.

7. For appropriate counter-

balancing, simply place a
CenSlide tube with equal volume
of urine or water opposite each
CenSlide tube placed in the rotor.

8. Seat the CenSlide tube firmly into

the rotor slot, pushing down on the cap until the CenSlide
tube can go no further. Close the lid.

When the lid is properly closed, the green “READY” light on
the control panel will stop flashing.

The lid must be "completely” closed, or the CenSlide 2000
Centrifuge will not operate.

Figure 2.2

Figure 2.3