Iris Sample Processing StatSpin® CenSlide 2000 Urine Sediment Analysis System User Manual

Page 9

Advertising
background image

Operation

OPERATOR’S MANUAL

CenSlide 2000 Urinalysis System

Page 2 - 1

This chapter covers the installation and general operation
of the CenSlide 2000 Urinalysis System.

2.0 Installing the Censlide 2000 Centrifuge

After removing the CenSlide
2000 Centrifuge from its box,
place it on a level working
surface close to an electrical
outlet. The CenSlide 2000
Centrifuge operates on 24 volts,
supplied by an external power
supply module. Plug the power
supply module into the electrical
outlet, and its power cord into the centrifuge’s power
receptacle located at the back of the centrifuge (Figure 2.1).
Mark a 300 mm clearance boundary around the centrifuge
to allow adequate ventilation and safety.

Please use only the power supply provided with the
equipment. To turn unit completely off, disconnect power
plug-in located at the rear of the instrument.

2.1 Safety Features

The CenSlide 2000 Centrifuge uses an electronically
controlled locking mechanism to prevent the centrifuge
from being:

1. Operated, except when the lid is completely closed

and latched; or,

2. Opened, while the centrifuge is in operation.

Figure 2.1

IMPORTANT: Do not attempt to defeat any safety
features of this equipment. If the equipment is not
used correctly, protection provided by the equipment
will be impaired.
To reduce the risk of injury -
• Do not operate with an unbalanced load.
• Inspect rotor for fatigue.

CAUTION: Incorrect tubes may damage rotor.

NO USER SERVICABLE PARTS -
Refer all service to qualified service personnal.

!