Trouble Shooting Guide

OPERATOR’S MANUAL

CenSlide 2000 Urinalysis System

Page 5 - 1

5.0 Trouble Shooting

OBSERVATIONS

ACTIONS

The CenSlide 2000 Centrifuge

Check that the power transformer

does not appear to have power.

is securely plugged into both the AC wall
outlet and the back of the centrifuge.

Try a different AC outlet to verify that
AC current is being supplied to the
power supply module.

Make sure the lid is securely closed and
locked in place.

Reset the CenSlide 2000 Centrifuge by
disconnecting and reconnecting the power
supply cord's plug to the centrifuge.

The CenSlide 2000

Balance appropriately by placing equally

Centrifuge appears to

filled CenSlide tubes opposite each

vibrate when operating.

other.

It does not appear that

Seat CenSlide tubes firmly and securely

the flicking mechanism

within the CenSlide 2000 Centrifuge

is functioning.

rotor.

Check that you are using the correct
power supply module. Use only the power
supply provided with the instrument.

Clean and dry the flicker button with alcohol
and a tissue or swab.