Trouble Shooting Guide

OPERATOR’S MANUAL

CenSlide 2000 Urinalysis System

Page 5 - 2

5.0 Trouble Shooting

OBSERVATIONS

ACTIONS

Sediment appears to pack along

Seat CenSlide tubes firmly and securely

the sides of the viewing area.

within the CenSlide 2000 Centrifuge
rotor.

If the specimen is extremely thick or turbid,
dilute the sample with saline.

CenSlide tubes leak.

Verify that the caps are firmly and securely in
place on the CenSlide tubes.

The CenSlide tubes are not made for multiple
use. Use a non–liquid filler to match the
weight of 5 mL of fluid when using as a
balance tube.

The lid latch is stuck in the

If the "OPEN / ABORT" button does not release

closed position.

the lid after the cycle end alert, disconnect the
power cord at the back of the unit and then
reconnect it to do a "RESET".
Retry the "OPEN / ABORT" button.

Place the bottom end of a Cen-Slide tube
into the slot in the lid latch and press against
the latch.

Carefully clean and dry the latch assembly.

If the lid O–ring is causing the lid to stick,
clean it and the underside of the lid with mild
soap and water.