Trouble Shooting Guide

OPERATOR’S MANUAL

CenSlide 2000 Urinalysis System

Page 5 - 3

5.0 Trouble Shooting

OBSERVATIONS

ACTIONS

The lid to the centrifuge is open

Disconnect the power cord at the

but the green light is not flashing.

back of the unit and then reconnect it.
This will cause a power "RESET".

The yellow light will not go on

Check that there is power to the

when the start button is pressed.

centrifuge by observing the green light.

Check that the lid is securely latched.

Microscopic observation

Make sure the CenSlide tubes have at least

and focus difficulty.

a one minute settling time in the horizontal
position after centrifugation.

Assure that the CenSlide tubes are properly
placed in the CenSlide Holder.

Using the 10x Objective, focus the
microscope on the top or bottom of the
CenSlide tube surface and then slowly
adjust up or down until you find the
observation plane. Then switch to your 40x
objective. If it comes in contact with the
CenSlide tube, you will need to obtain a
long working distance objective through
your local distributor or directly through
StatSpin.