Baumer GNAMG User Manual

Page 19

Advertising
background image

Manual_GNAMG_CANopen_EN.doc

19/41

Baumer IVO GmbH & Co. KG

21.11.12

Villingen-Schwenningen, Germany

Object

Sub-Index

Name

Format

Access

Default

Save

Meaning

02h Communication

Parameters

U32

rw

=“load“ (0x6C6F6164) to load

03h Application

Parameters

U32

rw

=“load“ (0x6C6F6164) to load

04h Manufacturer

Specific Parameters

U32

rw

=“load“ (0x6C6F6164) to load

1014h Emergency COB-ID

U32

rw

80h + Node-

ID

yes COB-ID of the emergency object

1016h Consumer heartbeat

time

ARR

yes

00h Maximum Subindex

U8

ro

1h

01h Consumer heartbeat

time

U32

rw

10000h yes Bit0..15 Consumer Heartbeat time in ms

Bit16..23 Node-ID

1017h Producer Heartbeat

Time

U16

rw

0h yes Producer Heartbeat time in ms

1018h Identity Object

REC

ro

00h Maximum Subindex

U8

ro

4h

01h VendorID

U32

ro

ECh

Vendor ID specified by CiA

02h Product Code

U32

ro

32h

Product Code:

0x32 = GNAMG.x225xxx
0x33 = GNAMG.x215xxx
0x34 = GNAMG.x235xxx

03h Revision Number

U32

ro

Works-

defined

yes Product revision number

04h Serial Number

U32

ro

Works-

defined

yes Unique successive serial number

1029h Error behaviour

ARR

Error behaviour

00h Maximum Subindex

U8

ro

1h

01h Communication error U8

rw

1h yes 0h = go to pre-operational mode

1h = no change of mode
2h = go to stop mode
3h = Node reset

1800h Transmit PDO1

Parameter

REC

00h Maximum subindex

U8

ro

5h

01h COB-ID

U32

rw

180h+id

ja

PDO ID = 180h + Node-ID

02h PDO Type

U8

rw

FEh

ja

FEh=User defined, cyclic

05h EventTimer

U16

rw

203h

ja

Cycle time in ms

1A00h Transmit PDO1

Mapping

ARR

00h Maximum Subindex

U8

ro

2h

01h Content of PDO1

I32

ro

61100020h

Slope angle Slope Long , Y-axis

02h Content of PDO1

I32

61200020h

Slope angle Slope Lateral, X-axis

2100h Baud rate

U8

rw

2h

ja

Setting the baud rate must be followed by
saving operation in EEPROM and re-init.

0=10 kBit/s
1=20 kBit/s
2=50 kBit/s
3=100 kBit/s
4=125 kBit/s
5=250 kBit/s
6=500 kBit/s
7=800 kBit/s
8=1000 kBit/s

2101h Node-ID

U8

rw

1h yes Node ID available from 1..127

Setting the baud rate must be followed by a
saving operation in EEPROM and re-init.

2110h Manufacturer_

Options

U32

rw

8h yes Bit3 = 0 no BusOFF reset

1 if BusOFF there is a bus reset

Bit5 = 0 Heartbeat protocol active

1 Nodeguarding protocol active

2201h Statistics

REC

00h Maximum subindex

U8

ro

3h

01h Total of position

errors

U32

ro

yes

Advertising