Battery Tips

NiCd batteries have several characteristics that can affect

both their performance and operating life. The following

tips advise you how to take advantage of these

characteristics to get the most out of your probe’s

battery.

!

Although NiCd batteries are rated for operation in

temperatures from -20 °C to +65 °C (-4 °F to +140

°F), operating the probe in extreme temperatures will

reduce operating time significantly. The optimum

operating temperature range for these batteries is

+20 °C to +30 °C (+68 °F to +86 °F).

!

The battery in the HI

n

4457 probe does not require

periodic "deep discharges" to reverse the capacity-

depleting "memory effect" caused by repeated

shallow discharges; however, undercharging can

reduce battery capacity. Therefore, after the charging

procedure is complete, be sure that the battery is fully

charged before resuming field operation.

!

If the battery exhibits low terminal voltage during

charging, or if it appears unable to acquire or maintain

an appreciable charge, individual cells in the battery

may be shorted or damaged. If, for any reason, your

battery needs replacement, contact Holaday Customer

Service for assistance.