APPENDIX B

HI

n

n

4457 OPERATING PROTOCOLS

Introduction

The information in this appendix assumes that you have

purchased the optional HI

n

4413P Fiber Optic / RS232

Interface, and are capable of communicating directly with

the HI

n

4457 probe via computer. No system readout is

required when using this configuration.

Communication Protocol

Data Type:

RS-232 Serial

Data Mode:

Asynchronous

Word Length:

7 bit

Parity:

Odd

Stop Bits:

1

Data Rate:

9600 baud

Information Transfer Protocol

The HI

n

4457 operates as a Controller Mode device. This

probe only responds to commands from another device;

it transmits no data without receiving instructions to do

so.

Command Structure

A command to an HI

n

4457 probe consists of 1) a

command letter, followed by 2) possible parameters, 3)

terminated with a carriage return. When it completes the

command, the probe responds with a string consisting of

1) a start character (":"), 2) the command letter, followed

by 3) data, if required, and terminated with 4) a carriage

return. If the command does not require the probe to

return any data, the probe simply responds with the

command letter and a carriage return. If an error occurs,

the probe responds with an error code, as detailed in