HI

n

4457 Manual

Page — 19

to perform a normal read cycle on the composite axis

signal. This procedure is executed for all ranges. When

the processor receives all the zero-field values, it stores

them in a special register; these values are subtracted

from all subsequent measurements. Therefore, a probe

which is zeroed while it is not in a zero field environment

will give erroneous readings.