6.0 THEORY OF OPERATION

Introduction

This section discusses the theory of operation and the

functions of the HI

n

4457 Isotropic Magnetic Field Probe.

A high-level block diagram (Figure 6

n

1) is included to aid

the discussion. The objective is to provide information

that enhances user understanding of the design of this

probe.

System Theory

The HI-4457 Broadband Isotropic Magnetic Field Probe

uses three orthogonal sensing elements. Each of the

elements responds to the magnetic field component

aligned with the particular element. The output of the

three elements are summed as vector quantities and an

output proportional to the resultant field generated.

A microprocessor is contained in the instrumentation

housing located at the base of the probe shaft. The

microprocessor generates an output corrected for

variations in the sensor using lookup tables located in the

processor memory. The resultant value is transmitted in

an ASCII digital format through the optical cable to the

readout. The probe's self-contained power supply

employs a 3.6 VDC NiCd battery which provides up to 17

hours of continuous operation. Refer to Appendix B for

details of the ASCII output formats.

For specialized applications, you may use a computer

with an RS-232 serial port to communicate directly with

the HI-4457 using the optional HI-4413P RS-232

interface.