7.0 APPLICATION CONSIDERATIONS

Introduction

The following subsections contain information designed

to help you maximize the effectiveness of the HI

n

4457.

Out-of-Band Considerations

Although the HI

n

4457 is nominally rated for operation

from 10 MHz to 1 GHz, it may respond to signals both

above and below these frequencies. Such responses

must be taken into account when performing certain

operations, such as zeroing.

Resolution Limitations

Limitations in probe resolution may result in a non-zero

reading when the receiver is zeroed. If this occurs, it

does not necessarily mean either that there is a problem

with the receiver or that your readings are inaccurate.

Receiver linearity is specified as + 0.5 dB full scale: in

addition, the variance of the A/D converter is + 2 least

significant bits. When using the most sensitive range on

a given probe, these specifications create the possibility

that, under zero field conditions, the receiver may display

a non-zero value.

Probe Support Structures

It is very important to keep conductive objects away from

the HI

n

4457. Any such objects in the proximity of the

probe may distort the near field and compromise

measurement accuracy. If your application requires

measurements from a fixed position, always mount the

probe on a non-metallic platform, using non-metallic

screws.