HI

n

4457 Manual

Null

Send the ASCII null character. This is a

special command that can be used as the

initial command to the probe after it is

powered up. The probe responds with

“N”.

Probe Output

Command

HI

n

4457 Response

B

Bxx.xx, where xx.xx is the battery voltage.

D1

Dxx.xxuuu, the short form output.

xx.xx is the reading. The position of the

decimal point depends upon the range

setting of the probe.

uuu = units

_A_ = A/m, mW2 = mW/cm², _A2 =

[A/m]²

(underscore indicates a space character)

D2

Dxx.xxuuurrrobaaat, the long form output.

xx.xx = the reading, as described for D1.

uuu = units, as describe for D1.

rrr = recorder out value (A 3-digit ASCII

number from 0 to 255).

o = over range indicator ("N" = ok, "O"

= over range).

b = battery status ("N" = safe operating

level, "W" = warning level, "F" = fail

Archived 4/2/10