Rice Lake 120 Plus Digital Weight Indicator User Manual

Page 10

Advertising
background image

6

120Plus Digital Weight Indicator Technical Manual

6

Hold Display

Note: Hold Mode is
selected through
the Configuration
Menu.

Toggle Hold Mode
1. Wait for the standstill annunicator (

). The wait time will depend on a selectable

parameter.

2. Press the (

) key to freeze the display. The HOLD annunicator is lit on the display.

3. Press the (

) key again to return to the weighing mode.

Note: Only one print is allowed during Hold mode.

Average Hold Mode
1. Place the weighing sample on the weight indicator and wait until the sample has stabilized.

2. Press the (

) key to start dynamic weighing. During dynamic weighing, the current

average weight appears on the display and then the dynamic weighing result is displayed with
the symbol HOLD. The weight indicator calculates the mean value from a hundred weighing
operations within AVGTM seconds.

3. Press the (

) key again to go back to weighing mode. Or wait until the HOLDTM

minute.

Note: When the (

) key is pressed for the first time, the weight indicator does not wait for

a stable signal. It is possible to press the (

) key at the beginning. Then place a weighing

sample on the scale until the sample has stabilized (via setting MOTBAND). Start dynamic
weighing. During the dynamic weighing process, the current average weight appears on the
display and the HOLD symbol flashes. The scale calculates the mean value from a hundred
weighing operations within AVGTM seconds. Then the dynamic result is displayed with the
symbol HOLD.

Auto Hold Mode
1. While the net weight is within the zero “dead” band, the indicator shows the current weight.

2. Press the (

) or (

) key to clear any residual weight and return the scale to the

zero state.

3. Press the (

) key to enter the Auto Hold Mode. The display shows the current weight

and the READY message alternately and flashes the HOLD symbol, if the weight moves inside
than the zero “dead” band.
4. Place the item to be weighed on the scale.
5. Once the weight moves outside the zero “dead” band and the signal is stable, the indicator
begins to calculate a long term average (selectable time, AVGTM) that compensates for any
movement in the mass. During the selectable time the signal must be stable, if not, time starts
again. The instrument flashes the HOLD symbol and shows the current average value.
6. The HOLD symbol is displayed when the final sample weight is shown on the display.

7. Press the (

) key to force the sample to be re-calculated. Press any other key to go

back to the weighing mode.
8. Once the weight has returned to zero “dead” band, the cycle can be repeated without any
pressing any key.
Note: The definition of zero dead band is five minutes (5DD or 5 scale divisions)

7

Keyed Tare

Using the numeric keypad, enter the desired tare weight. remember the weight is one of
pounds/kilograms/grams or ounces, whatever was set previously. If the number is not
entered, the display indicates an ER TOUT and this is normal. The indicator then returns to a
weighing mode automatically. If entered correctly, the tare weight is stored in the indicator.

Note: The keyed tare must be >0. Press the (

) key or

(

) keys and the indicator is now ready for weighing. The display shifts to net weight and

shows NET on the display.

8

Toggle Units

Press the (

) key to switch between primary and secondary units. The units annunicator

to the right of the display is lit.

9

Print Ticket

1. Wait for the standstill annunicator (

).

2. Press the (

) key to send data to the serial port.

Number

Name

Basic Operations

Table 1-4. Weighing Mode Operations

Advertising