Setpoint Operation

35

7.0

Setpoint Operation

Figure 7-1. Setpoint Operation

The following steps describe the internal steps of setpoint operation. To configure setpoints, refer to Section 3.7 on
page 25.

1. Power on the Toggle Flip-Flop Q=1, The CPU output may be high or low depending on the Sn.CONTACT

selection. Also, the XRL gate becomes an inverter.

2. If load (may be gross, net, or display parameters depending on the KIND command) is bigger than

Sn.WAVL, the output of XRL gate is high, otherwise it is low.

3. The timer will begin counting if the load is greater than Sn.WVAL. If not, the timer will load from SW2.

4. The Output of AND gate will be high if the output of OR gate is high. (Depends on Sn.CONDITI

selection.) Otherwise, wait until standstill is high (If Sn.CONDITI = STABLE).

5. Output (Q) of the Toggle F-F changed from 1 to 0 and the COIL also changed state. (NO to NC, or NC to

NO.) And SW1 & SW2 switch over other side, respective changed Comparator BASE and Timer constant.

Comparetor

Time r

Q

Toggle F-F

AND

OR

Standstill

Sn.CONDITI

NORMAL

STABLE

1

0

Sn.CONTACT

-

+

Sn.WV

AL

Sn.WV

AL
x (1-Sx.Hyster)

1

Sn.TIMER

Sn.DELA

Y

1

Gross

NET

KIND

R

+5V

N.O

N.C

Consant Load

COIL

0V

120 Plus

CPU

XRL

Power O

n

SET

0

0

SW1

SW2

Display