9.1.2

Reporting Commands

Reporting commands send speciÞc information to the
serial port. The commands listed in Table 9-2 can be
used in either setup mode and normal mode.

9.1.3

Clear and Reset Commands

The following commands can be used to clear and
reset the

920i

:

PCLR

: Program clear. Erases the loaded user program

(setup mode only).

RS

: Reset system. Resets the indicator without

resetting the conÞguration.

RESETCONFIGURATION

: Restores all conÞguration

parameters to their default values (setup mode only).
The

RESETCONFIGURATION

function can also be

initiated by pressing the

Reset Config

softkey under

the VERSION menu.

NOTE:

All load cell calibration

settings are lost when the RESETCONFIGURATION
command is run.

9.1.4

Parameter Setting Commands

Parameter setting commands allow you to display or
change the current value for a particular conÞguration
parameter (Tables 9-3 through 9-11).

Current conÞguration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:

command<ENTER>

Most parameter values can be changed in setup mode
only; setpoint parameters listed in Table 9-6 on
page 76 can be changed when in normal weighing
mode.

Use the following command syntax when changing
parameter values:

command=value<ENTER>

, where

value

is either a number or a parameter value. Use no spaces
before or after the equal (=) sign. If you type an
incorrect command, the display reads

??

.

For example, to set the motion band parameter on
Scale #1 to 5 divisions, type the following:

SC.MOTBAND#1=5D<ENTER>

For parameters with selectable values, enter the
command and equal sign followed by a question
mark:

command=?<ENTER>

to see a list of those values.

The indicator must be in setup mode to use this
function.

NOTE:

Some parameters are valid only if other

parameters or parameter values are specified. See the
configuration menus in Section

3.2 on page

21 for

information about parameter dependencies. Restrictions
for front-panel configuration also apply to serial command
configuration.

Command

Function

DUMPALL

List all parameter values

SPDUMP

Print setpoint configuration

VERSION

Write

920i

software version

S#n

Write one frame of stream format to scale
channel n

HARDWARE

Lists option cards installed in slots 1–14.
See Section 10.1.2 on page 88 for more
information about using the HARDWARE
command.

XE

Returns a10-digit code representing any
error conditions currently shown on the
front panel. See Section 10.1.4 on page 89
for more information.

Table 9-2. Reporting Commands

Command

Description

Values

SC.GRADS#n

Graduations

1–9999999

SC.ZTRKBND#n

Zero track band

OFF, 0.5D, 1D, 3D

SC.ZRANGE#n

Zero range

1.9%, 100%

SC.MOTBAND#n

Motion band

1D, 2D, 3D, 5D, 10D, 20D, OFF

SC.SSTIME#n

Standstill time

1-65535

SC.OVRLOAD#n

Overload

FS+2%, FS+1D, FS+9D, FS

SC.DIGFLTR1#n
SC.DIGFLTR2#n
SC.DIGFLTR3#n

Digital filtering

1, 2, 4, 8, 16, 32, 64, 128, 256

SC.DFSENS#n

Digital filter cutout sensitivity

2OUT, 4OUT, 8OT, 16OUT, 32OUT, 64OUT, 128OUT

SC.DFTHRH#n

Digital filter cutout threshold

NONE, 2D, 5D, 10D, 20D, 50D, 100D, 200D, 250D

SC.RATLTRAP#n

Rattletrap filtering

OFF, ON

Table 9-3. SCALES Serial Commands