ACK NAK Timeout Value

This option specifies the amount of time the reader waits for an ACK character from
the host following label transmission. The selectable timeout range is 200 millisec-
onds to 15,000ms (15 seconds) in 200ms increments. A selection of 0 disables the
timeout.

To set this value:

1. Determine the desired setting in milliseconds.

2. Divide the desired setting by 200 (setting is in 200ms increments). Pad the result

with leading zeroes to yield two digits. For example: 0 = 00, 5 = 05, 20 = 20, etc.

3. Scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Program-

ming Mode.

4. Scan the barcode: SELECT ACK NAK TIMEOUT VALUE SETTING.

5. Scan the appropriate two digits from the keypad in

Appendix E, Keypad

, that rep-

resent the duration which was determined in the steps above. You will hear a two-
beep indication after the last character.

6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Program-

ming Mode.

This completes the procedure. See

Table 10

for some examples of how to set this

feature.

Table 10. ACK NAK Timeout Value Setting Examples

NOTE

If you make a mistake before the last character, scan the CANCEL barcode to
abort and not save the entry string. You can then start again at the beginning.

STEP

ACTION

EXAMPLES

1

Desired Setting

200ms

1,000ms (1 sec.) 5200ms (5.2 sec.)

15,000ms (1 sec.)

2

Divide by 200

01

05

26

75

3

Scan ENTER/EXIT PROGRAMMING MODE

4

Scan SELECT ACK NAK TIMEOUT VALUE SETTING

5

Scan Two Characters From

Appendix E, Keypad

‘0’ and ‘1’

‘0’ and ‘5’

‘2’ and ‘6’

‘7’ and ‘5’

7

Scan ENTER/EXIT PROGRAMMING MODE