11/32

IPC-WP Series Service Manual

2.4 Memory

Switch

Operation

Display

1. C mode / F mode status

2. Press the Memory Switch.


2.5

C1 Mode – Country No. Setting

2.5.1 Country

No.

Table

Country No.

Item

Data

Wei-

ght

Digit

Bit

1 2 3 4 5 6 7

JPN ASIA

USA

AUS
CAN

EU

0 0 0 0 0 0 0

Start

range

0: ±10%
1: ±2%

1

10

0

D0

0 0 0 0 1 0 0

Zero

point

mark

0: Lights on at true zero
1: Lights on at
Provisional zero

2 D1

1 0 0 0 1 0 0

Below true zero
indication

0: "___"
1: Negative value

4 D2

0 0 0 0 1 0 0

Cleaning tare weight
by pressing ZERO key

0: No
1: Yes

8 D3

0 0 1 0 0 0 1

Decimal point
Indicator

0: "."
1: ","

1

10

1

D4

0 0 0 1 0 0 0

Over-scale

indication

0: BLANK
1: "OL"

2 D5

0 0 0 0 1 0 0

Tare

subtraction

0 : Not possible
1: Possible

4 D6

0 0 0 0 0 1 1

Tare clear with ZERO
key

0

: Possible

1

: Not Possible

8 D7

0 0 0 0 0 0 0

Stabilized,
re-stabilized
frequency

0: 5 times.
1: 8 times.

1

10

2

D8

0 0 0 0 0 0 0

Re-stabilization
Starting range

0: 2 times over
1: 4 times over

2 D9

0 0 0 0 0 0 0

Stabilized/re-
stabilized range

0: 3 times
1: 5 times

4 D10

1 0 0 0 0 0 0

Section

adjustment

0: No
1: Yes (JAPAN)

8 D11

804 000 010 020 04E 080 090 ←Display of F mode Measurement Condition