IPC-WP Series Service Manual

6. Remove the Approval seal.

7. Remove the 10 M4 screws located around the

base.

Caution

When re-assembling the scale, first fasten the
screws tentatively, then tighten them completely
one by one in a diagonal sequence.
If you strongly tighten a particular point first, the
water proofing property will be deteriorated
because you cannot push in the packing between
the upper case and the base evenly.

8. Pull out the connector from CN1 on the main

board, then remove the upper case.


Perform this procedure in reverse for assembly.


Caution

Affix a new Sheet protect.
When affixing the new sheet protect, completely
remove the starch left on the bracket.
Otherwise, the water proofing property will be
deteriorated.。

Sheet

(Approval seal)

O-ring

M4 Screw

Base

CN1