Table 8-1. Troubleshooting Chart

Condition

Cause

No indicator display

• Power cable disconnected
• Power connection fault
• Defective indicator

Erratic indicator readings

• Erratic supply voltage - loose cables/connectors
• Defective cable or connectors
• Defective load cell/s
• Deffective indicator
• Loose load cell bolts

Incorrect weight readings
(not erratic)

• Incorrect calibration
• Dirty connectors
• Defective cable
• Defective load cells
• Top load cell mounting bolts too long - always weighing light
• Body rubbing on the chassis
• Build-up of dirt or grit in the loadcell slot

Unable to calibrate

• Top load cell mounting bolts too long to calibrate maximum weight

Unable to zero

• Defective load cell
• Loose or broken connectors

Note

Appendix

27

Your RLWS LoadRunner system should be serviced & re-calibrated annually.