Page 20 MSI-9300 Crane Scale™

User Guide

MSI-9300 Porta-Weigh+ Crane Scale

SECTION 4 – THE EMBEDDED CELLSCALE

For a complete guide to all the features of the embedded CellScale, refer to the MSI-9000 User Guide.

ACCESSING THE CELLSCALE

MSI-9300 Series crane scales are a shell interface to an embedded CellScale. To access the advanced features

of the CellScale, a Comm Port is provided on the front casting of the crane scale. This Comm Port is only used

for interfacing to the embedded CellScale and cannot be used as an output of any kind.
The CellScale offers Set Points, ID Codes, and other advanced features which are controlled either through a

remote Indicators such as the MSI-3750CS, the MSI-9850, or the MSI-9750A Handheld. These features are

also accessed by connecting a terminal program in a PC to the Comm Port. Tested terminal programs that

function well with the CellScale include ProComm Plus, Qmodem, HyperTerm (shipped with many versions

of Windows), WinTerm, and ZTerm (for Macintosh). Usually a VT102 emulation is used. All commands are

single key except when entering data.
Settings for your terminal program are: 9600 baud, 8 bit, 1 start bit, 1 stop bit, no parity, Handshaking usually

not required, but you can turn on Xon/Xoff if you desire.
While the Comm Port is in use, the display of the MSI-9300 Series crane scale is in standby with the message

“C-Port” or “H-Port” displayed. The only key that will function in this mode is the Power key.
The CellScale Terminal Access Mode is used to set Network parameters, Scale parameters, Set Points, and

other various setup conditions. Refer to the MSI-9000 CellScale User Guide for detailed information. The Ter-

minal Access Mode is used for calibration on MSI-9300 Series crane scales that do not have the RF modem.

CONNECTION PROCEDURE

The MSI-9300 must be on and finished with its startup sequence before connecting the

terminal cable to a PC.

1) Without connecting the PC or Terminal to the MSI-9300, turn the MSI-9300 on. Wait for the startup

sequence to complete (usually about 10-15 seconds). When weight is displayed, the MSI-9300 is ready.

2) Press and Hold the

NET/GROSS

key and then push

TEST

. The display should change to “H-PORT”.

3) Connect the PC to the MSI-9300 through the Comm Port. Start your terminal emulation program with set-

tings 9600-N-8-1

4) You can now access the Terminal menus in the CellScale. Push any key to refresh the menu.
5) Once in the Terminal Access Mode menus, commands are usually a single number or letter. Refer to the

9000 CellScale User Guide for further information.

COMM PORT CABLE

The MSI-9300 comes standard with one Comm port cable wired for RS-232 (MSI P/N 501705-0001) follow-

ing the AT standard for 9 pin serial cables (DCE). An unterminated cable is available (MSI P/N 12023) if you

wish to wire your own serial cable for RS-232.

1

2

3

4

5

6
7

8

(NC) 9

Brown (TD)

Black (RD)

Blue (GND)

Grey (CTS)

White (RTS)

Jumper 1-6-4

RS-232 DCE

9-Pin Female ‘D’

Solder Cup View

Standard wiring for direct

9000 to Computer connection

Comm Port Cable Color Code RS-232
Brown – Transmit output from CellScale, connect to

receive of DTE.

Black – Receive input to CellScale, connect to trans-

mit of DTE.

Grey – CTS Input to CellScale. Connect to RTS or

RTR output of DTE.

White – RTS/RTR output from CellScale. Connect to

CTS input of DTE.

Blue – Signal Ground
Drain Wire – Connect to metal shell.