Page 26 MSI-9300 Crane Scale™

User Guide

MSI-9300 Porta-Weigh+ Crane Scale

***** Channel Calibration *****

Calibrating Ch # 1-1, named: Fred2...............
Precision in Kilograms : 10000. x 2.

Test weight # 0 in Kilograms : 0.
Test weight # 1 in Kilograms : 5000.
Test weight # 2 in Kilograms : 2000.

Load the scale with test weight # 2
Press (Y) when scale is loaded and steady
Press (B) to backup
Press (ESC) to complete and store calibration

Raw A/D counts: 4528050

17) Ensure the 2nd test weight is on the scale and steady. When it is, press [Y].

***** Channel Calibration *****

Calibrating Ch # 1-1, named: Fred2...............
Precision in Kilograms : 10000. x 2.

Test weight # 0 in Kilograms : 0.
Test weight # 1 in Kilograms : 2000.
Test weight # 2 in Kilograms : 5000.

Test weight # 3 in Kilograms : 2000.
Press (Y) if this is correct
Press (N) to change
Press (B) to backup
Press (ESC) to complete and store calibration

18) After the second span point calibration, the CellScale rearranges the test weights in ascending order. The

following menu appears for cal point 3. This is the end of our example calibration. Push [ESC] to end cal,

store constants, and return to the main menu. If more span points are desired, repeat steps 16 and 17 up to

a maximum of 9 span points.

CELLSCALE SETTINGS FOR MSI-9300

Certain Terminal Access Settings are required for compatibility with the MSI-9300 display board. These set-

tings should not be changed. The default settings should be:
MAIN > SYSTEM > POWER: Set to AC (Display board checks battery, do not set to 12V SLA).
MAIN > SYSTEM > LEDs: Turn all off and turn Power to Low. (LEDs can’t be seen anyway, saves power to

turn them off).
Do not modify Comm Port 1 baud rate settings. This could interfere with communications with the display

board.

RF SYSTEM SETUP

MSI-9300 Series crane scales use an embedded CellScale to communicate through an advanced 2.4GHz

Spread Spectrum modem. Details of the RF Network setup is covered in the MSI-9000 CellScale User Guide.

LEGAL STANDARD

The MSI-9300 Series scales must be configured properly for use in Legal -for-Trade transactions. For most

industrial applications, LFT settings are not required. The Industrial and Metric settings have the same accu-

racy performance as the NTEP and OIML settings. Unless you are required by law to use the scale in LFT