Advanced entry calibration (continued), Step operator action displayed message comment, Proceed according to your media type – Rice Lake Datamax M-Class User Manual

Page 82: Next, press the, Or, for reflective media

Advertising
background image

74

M-Class

Advanced Entry Calibration (continued)

Step

Operator Action

Displayed Message

Comment

11

Proceed according to your media
type:

Die-cut stock: strip the media
from the backing and then
reinsert it into the sensor; adjust
the Sensor Eye Mark over the
center of the backing.

Notched stock: adjust the
Sensor Eye Mark over the center
of the notch.

Reflective stock: Adjust the
Sensor Eye Mark over the center
of the facedown black mark.

Press the

Key to select

the setting and then record the
sensor reading in a table similar
to the one shown after Step 12,
as a TOF Value.

Next, press the

FWD

Key to

increment the Gain Number.

Repeat for each of the Gain
Numbers (00-32).

!

HBJO!USBO!!!!!=zzz?!

+11!!!!!!!!)1!.!42*!

!

Or, for reflective media:

HBJO!SFGM!!!!!=zzz?!

+11!!!!!!!!)1!.!42*!

Where ‘yyy’ is a numerical value representing the
current sensor reading: useful in locating the best
sensor position.

Note: Never position the sensor over a perforation

when recording a sensor reading.

Note: Do NOT move the position of the Media Sensor after it has been adjusted.