7.1.1.2.1. Calibration Adjustment

The Calibration Adjustment function allows the internal calibration mass value to be adjusted

with an external calibration weight that has a traceable class certification. To perform calibration

adjustments follow the steps below:

1. Press the “T” (YES) key when “CALIB” is displayed in the calibration setup menu.

2. The scale will display CAL, indicating that the calibration adjustment mode has been entered.

3. Before calibration adjustment can begin, the scale will perform automatic tare. TARING will be

displayed during auto tare.

4. Once the scale has been tared the current calibration mass setting will be displayed. To change the

calibration mass, press the 0 (NO) key. After making the selection press the “T” (Yes) key to

accept and confirm the setting.