Torbal AGC Series Manual User Manual

Page 62

Advertising
background image

Instruction Manual - TORBAL AGC Series

7. To change the OVER limit follow the directions in Step 4 of this section and Table 8.C.4.

8. Once both limits have been set the scale will display a TARE command, indicating that the scale

must be tared. Before using the function place a container on the pan and press “T” to tare.

9. The scale is now ready to use. To perform check weighing place your object on the pan to check

whether the weight of the object falls within the designated range. If the weight is less than the

lower limit the scale will display the weight of the object and the command UNDER. If the weight

falls within the range, the ACCEPT command will be displayed. If the weight exceeds the upper

limit, the scale will display OVER.

Note: To change the limit setting press “C” and repeat Steps 1 to 7. If the same limits are used,

the scale can be tared without clearing the function and the new object can be weight-checked.

To exit check weighing and switch to a different function, press “C” to clear the current function

and then press “Function” to select a different function (follow Steps 1 through 3 from Chapter 8

on page 44).

62

Advertising