3. After confirming the sample size selection, the display will read SAPL On, indicating that you

may now place the sample on the pan (10 pieces if SPL-10 is selected).

4. Manually count out the sample and place it in the counting container located on the pan.

5. The weight of the sample will be displayed as it is placed on the pan.

Note: Minimum individual piece weight must be equal to or greater than 3e. Attempting to set an

average individual piece weight which is lower than 3e will result in s Sample Low Error (SAPL

LO). To see this limit expressed in grams turn to detailed specifications on page 4.