Calibration report – Torbal DRX300 Manual User Manual

Page 24

Advertising
background image

Instruction Manual - TORBAL DRX 5 Series

4. In order to confirm internal calibration and proceed with calibration press the calibration key

again.

1.4.1.Calibration Report

After calibration has been performed, a calibration report can be printed. The calibration report includes data of the
current internal calibration weight. To print the calibration report, follow these steps:

1. Press the MENU key to enter the main menu.

2. Use the navigation keys to select Calibration and press ENTER

3. Use the navigation keys to select Report Printout and press ENTER

4. The calibration report will be printed to a printer or transferred to a PC via the RS232 port.

24