3 Technical data

14

Melting Point M-565 Operation Manual, Version D

3 .2

Technical data overview

Table 3-3: Technical data of the Melting Point M-565

Melting Point M-565

Manual melting point determination

Manual boiling point determination

Automatic melting point determination

Automatic boiling point determination

Homogeneous sample loading

Positions for melting capillaries

3

Positions for boiling capillaries

1

Precision magnifying lens

Magnification of lens

2.5×

Digital camera

Video function

Magnification, display

6 x

Display

Colour, TFT, 320×240, 3.5˝

Determination temperature range

Ambient + 10 °C to 400 °C

Temperature resolution

0.1 °C

Accuracy melting point at 0.5 °C/min
up to 250 °C

± 0,3 °C

250 °C to 400 °C

± 0,3 °C to ± 0,5 °C

Repeatability melting point at 0.5 °C/min

± 0.1 °C

Accuracy boiling point at 1.0 °C/min up to
400 °C

± 0,5 °C

Repeatability b oiling point at 1.0 °C/min

± 0.3 °C

Temperature gradients, °C/min

0.1, 0.2, 0.5, 1, 2, 3, 5, 10, 20

Heat-up time (50 °C–350 °C) at 25 °C

~ 4 min

Cool-down time (350 °C–50 °C) at 25 °C

~ 13 min

Electrical supply

100–240 V (±10%), 50–60 Hz

Power consumption

150 W

Contact termination

L, N, PE

Video run time

350 min at 1 °C/min, 700 min at 0.5 °C/min

Approval

CE, CSA, UL

Dimensions (W×H×D), mm

190Ч200Ч370

Weight, kg

4.5

Environmental conditions

For indoor use only

Temperature

5–40 °C

Altitude

up to 2000 m a.s.l.

Humidity

Maximum relative humidity 80% for temperatures up to 31 °C,
decreasing linearly to 50% relative humidity at 40 °C

Over voltage category

II

Degree of protection

IP 20

Pollution degree

2

Advertising