8 Troubleshooting

53

Melting Point M-565 Operation Manual, Version D

Table 8-1: General malfunctions and their remedy

Malfunction

Possible cause

Remedy

Sample decomposes during the
melting process (formation of
bubbles, sample turns brown, etc.)
Apparatus parameters:
Apparatus is not or is poorly cali-
brated

Regular calibration of apparatus with
BUCHI calibrating substances

Method “According to pharmaco-
poeia” or “thermodynamic melting
point determination” is incorrectly
selected

Alter the corresponding parameters

No results

Setting point is too close to the
melting point

Select setting point 5–10 °C below
melting point

No statistic

Setting parameter Statistic is
disabled

Enable the Statistic in the Settings

Instrument does not heat

Heating defective

Contact BUCHI customer service.

Table 8-2: Malfunctions with Melting Point Monitor software

Malfunction

Possible cause

Remedy

No USB connection

USB cable is not connected or defec-
tive.

Check USB connection, replace if
necessary.

Driver not found

Device connected for the first time.

Install recommended driver.

Table 8-3: Warning messages

Warning number

Possible cause

Remedy

Warning 01

Calibration outdated.

Calibrate or verify device.

Warning 02

Housing temperature too high.

Cool down device and check clearance of vent
holes and fans.
Check ambient temperature.

Warning 03

Device not calibrated.

Calibrate device.

Table 8–4: Error messages

Error number

Possible cause

Remedy

Error 01

Memory data loss, all data are reset.

Calibrate device.
In case of recurrence contact customer service.

Error 02

Automatic restart, firmware problem possible.

In case of recurrence contact customer service.

Error 03

Board temperature sensor defective.

Contact customer service.

Error 04

Temperature sensor defective.

Contact customer service.

Error 05

Heating defective.

Contact customer service.

Error 06

Cooling fan defective.

Contact customer service.

Error 07

Housing fan defective.

Contact customer service.

Error 08

Camera defective, communication loss.

Contact customer service.

Error 09

24 V fuse defective.

Contact customer service.

Error 10

24 V input voltage missing.

Contact customer service.

Advertising