45

Melting Point M-565 Operation Manual, Version D

6 .4 .4

Verification

To verify the calibration, proceed as follows:
• Measure all standards (BUCHI recommends using the verification set (11055019) but it is also

possible to use your internal standards.

NOTE
The verification is not a guided process.

• When all measuring results were within the required tolerances, click on

Verified. The following

screen appears:

• Enter the following password:

VER. The current date is now indicated under “Last verification”.

6 .5

Settings, SysInfo, Test

Test: Functional test of the instrument. To per-
form the test of several functions follow carefully
the instuction on the screen. If a function test
shows not

ok, please contact the customer

service. The Test protocol is described in sec-
tion 6.5.1.
SysInfo: Instrument gives information regarding
settings and connected devices.
Settings: This menu can be used to change
parameters. The SysInfo protocol is described
in section 6.5.2.