3 creating a method, 4 adjusting parameters during a determination – BUCHI Melting Point M-565 User Manual

Page 40

Advertising
background image

6 Operation

40

Melting Point M-565 Operation Manual, Version D

73.2

76.4

73.4

76.6

Delay: to ensure equilibrium
between oven temperature
and sample temperature.
During this delay time some
air bubbles turns out of the
boiling point capillary. This
request can be changed in
the setting “Boiling sample
conditioning” (see 6.5).

• To run through the results,

turn the rotary knob up and
down.

Video replay is not available
for boiling point determination.

• Press

End or STOP to

return to the idle screen.
• Press

START to measure

again with the identical
measuring parameters.

6 .3 .3

Creating a method

NOTE
The process of creating a boiling point method is identical that used to create a melting point method
(see 6.2.3).

6 .3 .4

Adjusting parameters during a determination

During preheating or a determination, measuring parameters can be modified. This function is possible
only if enabled by the setting “Param. adj. during meas.” (see 6.5).

a

• Turning the rotary knob

displays the current param-
eter set.

• Adjust the parameters by

pressing Edit. Adjusting the
parameters does not affect
the saved method.

If the temperature gradient
is changed during a mea-
surement, the result shows
CHANGED (a) for the tem-
perature gradient.