8 troubleshooting, 1 malfunctions and their remedy, Malfunctions and their remedy – BUCHI Melting Point M-565 User Manual

Page 52: 8troubleshooting, 8 .1 malfunctions and their remedy

Advertising
background image

8 Troubleshooting

52

Melting Point M-565 Operation Manual, Version D

8

Troubleshooting

The following section describes how to resume operation of the instrument in the event of any minor
problem. It lists some possible occurrences, their probable cause, and suggests how to remedy
the problem. The troubleshooting table below lists possible instrument malfunctions and errors and
describes courses of action that can be used by operators to correct some of those problems. The
appropriate course of action is listed in the column “Remedy”.
More complicated malfunctions or errors are usually handled by a BUCHI technical engineer who has
access to the official service manuals. In this case, please contact your local BUCHI customer service
agent.

8 .1

Malfunctions and their remedy

Table 8-1: General malfunctions and their remedy

Malfunction

Possible cause

Remedy

Instrument does not work

Main switch off

Switch on mains switch

Instrument is not connected to mains
supply

Check mains connection

No or unreadable printout

Not activated in the settings

Activate automatic printout in the
settings

Bad cable connection

Check cable connection

Printer switched OFF

Switch ON printer

No paper

Replace paper roll

Baud rate settings of printer faulty

See section 8.1.1

No melting curve is shown

Not activated in the settings

Activate graphic in the settings

No automatic detection of samples

Manually register the melting point/
boiling point

Measuring finished before detecting
a result for all samples

Turn off in the settings “Stop after
detection” and check stop tempera-
ture.

In the automatic melting point deter-
mination, no value and no melting
curve are determined

Sample is unstable, it decomposes,
turns brown, or melts non-uniformly

The temperature parameters have
been entered incorrectly and the
sample does not melt at all

Automatic detection is disabled

e.g. select higher maximum point

Enable the Automatic detection in the
Settings

Sometimes no results or only 1 or 2
results instead of 3 results

START button was pressed before
insertion of all sample capillaries

Put all sample capillaries in the oven
first, then press the START button.
Do not move the sample capillaries
again.

Unexpected results

Sample-specific preparations:
Sample is not dry or is contaminated
by another substance

Sample should be dried before use